title: Fix the Intro.ini Missing or Incorrect Error on Your Tecan EVOware
description: Fix the “C:\ProgramData\Tecan\EVOware\Intro.ini is missing or incorrect” error on your Tecan EVOware with step-by-step troubleshooting tips.
image: https://blog.seqgen.com/hubfs/tecan-intro-is-missing.png
---

- [MARKETPLACE](https://seqgen.com/marketplace/)
- [CAREERS](https://seqgen.com/careers/)
- [TERMS & CONDITIONS](https://seqgen.com/wp-content/uploads/2026/03/POL_011-Rev.A_Terms-and-Conditions-of-Sales-for-Products-and-Services.pdf)
- [ISO 9001 CERTIFICATE](https://seqgen.com/wp-content/uploads/2026/01/SeqGen-Final-Cert_2026-2029.pdf)
- [SALES INQUIRY](https://seqgen.com/sales-inquiry/)
- [SERVICE REQUEST](http://seqgen.com/service-request/)
- [TESTIMONIALS](https://seqgen.com/testimonials/)

[![SeqGen](https://seqgen.com/wp-content/uploads/2020/03/seqgen_logo.png)](https://seqgen.com/)

- [SeqMo Rewards](https://seqgen.com/seqmo-rewards/)
- [HOME](https://seqgen.com/)
- [ABOUT US](https://seqgen.com/about-us/) 
    - [Our History](https://seqgen.com/about-us/our-history-2/)
    - [Our Engineers](https://seqgen.com/about-us/our-engineers/)
- INSTRUMENT MODEL 
    - [LC/MS](https://seqgen.com/instrument-model/lc-ms/) 
          - [Sciex 3000 system](https://seqgen.com/instrument-model/lc-ms/api-3000/)
          - [Sciex 3200 system](https://seqgen.com/instrument-model/lc-ms/api-3200/)
          - [Sciex 4000 system](https://seqgen.com/instrument-model/lc-ms/api-4000/)
          - [Sciex 4500 system](https://seqgen.com/instrument-model/lc-ms/api-4500/)
          - [Sciex 5500 system](https://seqgen.com/instrument-model/lc-ms/api-5500/)
          - [Sciex 6500 system](https://seqgen.com/instrument-model/lc-ms/api-6500/)
          - [LC systems](https://seqgen.com/instrument-model/lc-ms/lc-systems/)
    - [DNA Analyzer](https://seqgen.com/instrument-model/dna-analyzer/) 
          - [ABI 3100](https://seqgen.com/instrument-model/dna-analyzer/abi-3100/)
          - [ABI 3130/3130XL](https://seqgen.com/instrument-model/dna-analyzer/abi-3130-3130xl/)
          - [ABI 3500/3500XL](https://seqgen.com/instrument-model/dna-analyzer/abi-3500-3500xl/)
          - [ABI 3730/3730XL](https://seqgen.com/instrument-model/dna-analyzer/abi-3730-3500xl/)
    - [Real-Time PCR](https://seqgen.com/instrument-model/real-time-pcr/) 
          - [ABI 7300/7500](https://seqgen.com/instrument-model/real-time-pcr/abi-7300-7500/)
          - [ABI 7900](https://seqgen.com/instrument-model/real-time-pcr/abi-7900/)
    - [Thermal Cyclers](https://seqgen.com/instrument-model/thermal-cyclers/) 
          - [ABI 2720](https://seqgen.com/instrument-model/thermal-cyclers/abi-2720-2/)
          - [ABI 9700](https://seqgen.com/instrument-model/thermal-cyclers/abi-9700/)
          - [ABI 9800](https://seqgen.com/instrument-model/thermal-cyclers/abi-9800/)
    - [Liquid Handler](https://seqgen.com/instrument-model/liquid-handler/) 
          - [Tecan Freedom Evo](https://seqgen.com/instrument-model/liquid-handler/tecan-freedom-evo/)
    - [Microplate Reader](https://seqgen.com/instrument-model/microplate-reader/) 
          - [Coming Soon](https://seqgen.com/instrument-models/microplate-reader/)
- [SALES](https://seqgen.com/sales/)
- [SERVICE](https://seqgen.com/service/)
- [CONTACT](https://seqgen.com/contact/)
- [BLOG](https://blog.seqgen.com)

- [HOME](http://seqgen.com/)
- [ABOUT US](https://seqgen.com/about-us/) 
    - [Our History](https://seqgen.com/about-us/our-history-2/)
    - [Our Engineers](https://seqgen.com/about-us/our-engineers/)
- INSTRUMENT MODEL 
    - [LC/MS](https://seqgen.com/instrument-model/lc-ms/) 
          - [Sciex 3000 system](https://seqgen.com/instrument-model/lc-ms/api-3000/)
          - [Sciex 3200 system](https://seqgen.com/instrument-model/lc-ms/api-3200/)
          - [Sciex 4000 system](https://seqgen.com/instrument-model/lc-ms/api-4000/)
          - [Sciex 4500 system](https://seqgen.com/instrument-model/lc-ms/api-4500/)
          - [Sciex 5500 system](https://seqgen.com/instrument-model/lc-ms/api-5500/)
          - [Sciex 6500 system](https://seqgen.com/instrument-model/lc-ms/api-6500/)
    - [DNA ANALYZER](https://seqgen.com/instrument-model/dna-analyzer/) 
          - [ABI 3100](https://seqgen.com/instrument-model/dna-analyzer/abi-3100/)
          - [ABI 3130/3130XL](https://seqgen.com/instrument-model/dna-analyzer/abi-3130-3130xl/)
          - [ABI 3500/3500XL](https://seqgen.com/instrument-model/dna-analyzer/abi-3500-3500xl/)
          - [ABI 3730/3730XL](https://seqgen.com/instrument-model/dna-analyzer/abi-3730-3500xl/)
    - [REAL-TIME PCR](https://seqgen.com/instrument-model/real-time-pcr/) 
          - [ABI 7300/7500](https://seqgen.com/instrument-model/real-time-pcr/abi-7300-7500/)
          - [ABI 7900](https://seqgen.com/instrument-model/real-time-pcr/abi-7900/)
    - [THERMAL CYCLERS](https://seqgen.com/instrument-model/thermal-cyclers/) 
          - [ABI 2720](https://seqgen.com/instrument-model/thermal-cyclers/abi-2720-2/)
          - [ABI 9700](https://seqgen.com/instrument-model/thermal-cyclers/abi-9700/)
          - [ABI 9800](https://seqgen.com/instrument-model/thermal-cyclers/abi-9800/)
    - [LIQUID HANDLER](https://seqgen.com/instrument-model/liquid-handler/) 
          - [Tecan Freedom Evo](https://seqgen.com/instrument-model/liquid-handler/tecan-freedom-evo/)
- [SALES](https://seqgen.com/sales/)
- [SERVICE](https://seqgen.com/service/)
- [CONTACT](https://seqgen.com/contact/)
- [SERVICE REQUEST](https://seqgen.com/service-request/)
- [SALES INQUIRY](https://seqgen.com/sales-inquiry/)
- [BLOG](https://blog.seqgen.com/)
- [TESTIMONIALS](https://seqgen.com/testimonials/)
- [ISO 9001 CERTIFICATE](https://seqgen.com/wp-content/uploads/2026/01/SeqGen-Final-Cert_2026-2029.pdf)

# Fix the Intro.ini Missing or Incorrect Error on Your Tecan EVOware

10 September 2025

[Tecan Freedom Evo](https://blog.seqgen.com/topic/tecan-freedom-evo)

[0](https://blog.seqgen.com/fix-the-intro.ini-missing-or-incorrect-error-on-your-tecan-evoware#comments-listing)

![Fix the Intro.ini missing or incorrect error on your Tecan EVOware](https://blog.seqgen.com/hs-fs/hubfs/tecan-intro-is-missing-2.1.png?width=2816&height=1237&name=tecan-intro-is-missing-2.1.png)

Are you seeing the frustrating error message "C:\\ProgramData\\Tecan\\EVOware\\Intro.ini is missing or incorrect" when you power on your instrument? This issue can be caused by an improper shutdown of the software or the system, which can corrupt or delete critical configuration user record files.

![tecan-intro-is-missing](https://blog.seqgen.com/hs-fs/hubfs/tecan-intro-is-missing.png?width=512&height=270&name=tecan-intro-is-missing.png)

The good news is, you can resolve this problem with a simple repair process. This blog post will guide you through the steps to get your instrument up and running again.

## Understanding the Problem

The Intro.ini file is a crucial configuration user record file for the EVOware software. When this file is missing or corrupted, the software cannot start correctly, leading to the error message you're seeing. This typically happens if the computer is turned off before the software has properly closed all its files.

## The Solution: Repairing the Intro.ini File

The most reliable way to fix this issue is to force the EVOware installation software to generate a new, clean Intro.ini file. Here’s how you do it:

Step 1: Backup the Existing Intro.ini File

Before you do anything else, it's a good practice to back up the file that's causing the problem. This might seem counterintuitive since the file is reported as "missing or incorrect," but it ensures you have a backup copy in case something goes wrong.

- Navigate to the following directory: C:\\ProgramData\\Tecan\\EVOware\\
- Locate the file named Intro.ini.
- Right-click on the file and select **"Copy"**
- Paste the file into a different folder, like your desktop or a new folder you create for this purpose.

Step 2: Delete the Original Intro.ini File

Now that you have a backup, you can safely remove the problematic file from its original location.

- Go back to the C:\\ProgramData\\Tecan\\EVOware\\ folder.
- Right-click on the Intro.ini file.
- Select **"Delete"**

This action will ensure that when you run the repair function, the software detects that the file is gone and will be prompted to create a new one.

Step 3: Run the EVOware Repair Function

The next step is to use your EVOware installation software to initiate a repair.

- Insert your EVOware installation media (CD/DVD) or locate the installation file(.exe) on your computer.
- Run the installer.
- Instead of seeing the options to "Install" or "Uninstall," you should see a **"Repair"** option. Select it.
- Follow the on-screen prompts to complete the repair process. The software will scan for missing or corrupted files and replace them.

Step 4: Verify the Fix

Once the repair is complete, it's time to check if the issue is resolved.

- Power on your instrument and launch the EVOware software.
- The software should now be able to start correctly without displaying the "Intro.ini is missing or incorrect" error.

If the error persists after following these steps, you may need to perform a full reinstallation of the software. However, the repair process is often all that's needed to fix this specific issue.

By following this simple troubleshooting guide, you can quickly resolve a common error and get back to using your instrument without further interruption.

If the error persists after these steps, don’t hesitate to [contact our service team](https://seqgen.com/service-request/) for expert assistance.

---

 

SeqGen specializes in the repair, maintenance, and refurbishment of essential laboratory instruments used in DNA sequencing and genetic research. Our expertise covers a wide range of equipment, including DNA sequencers like the [ABI 3730/3730XL](https://seqgen.com/instrument-model/dna-analyzer/abi-3730-3730xl/), [3500/3500XL](https://seqgen.com/instrument-model/dna-analyzer/abi-3500-3500xl/), SCIEX LC/MS mass spectrometers such as the [SCIEX 6500+](https://seqgen.com/instrument-model/lc-ms/api-6500/), [TECAN Freedom EVO](https://seqgen.com/instrument-model/liquid-handler/tecan-freedom-evo/) liquid handlers, real-time PCR systems, thermal cyclers, and microplate readers. With experience in both current and legacy models, SeqGen provides cost-effective solutions at significantly lower prices than leading competitors.

### Written by [Service Team](https://blog.seqgen.com/author/service-team)

Search this site on Google

Search Google

### Subscribe to Email Updates

## [![New Call-to-action](https://hubspot-no-cache-na2-prod.s3.amazonaws.com/cta/default/1939955/32eb2ce8-8e54-4e61-ad50-57f6b9b23d18.png)](https://hubspot-cta-redirect-na2-prod.s3.amazonaws.com/cta/redirect/1939955/32eb2ce8-8e54-4e61-ad50-57f6b9b23d18) 

### Recent Posts

### Posts by Topic

- [DNA Analyzer (49)](https://blog.seqgen.com/topic/dna-analyzer)
- [Others (42)](https://blog.seqgen.com/topic/others)
- [Crossword Puzzle (11)](https://blog.seqgen.com/topic/crossword-puzzle)
- [3730 Series Genetic Analyzer (10)](https://blog.seqgen.com/topic/3730-series-genetic-analyzer)
- [Tecan Freedom Evo (9)](https://blog.seqgen.com/topic/tecan-freedom-evo)
- [AI 101 (6)](https://blog.seqgen.com/topic/ai-101)
- [Archive (6)](https://blog.seqgen.com/topic/archive)
- [Events/News (6)](https://blog.seqgen.com/topic/events-news)
- [3500 Series Genetic Analyzer (5)](https://blog.seqgen.com/topic/3500-series-genetic-analyzer)
- [LC-MS (5)](https://blog.seqgen.com/topic/lc-ms)
- [Real-Time PCR (4)](https://blog.seqgen.com/topic/real-time-pcr)
- [3130 Series Genetic Analyzer (2)](https://blog.seqgen.com/topic/3130-series-genetic-analyzer)

see all

### Archives

- [July 2026 (1)](https://blog.seqgen.com/archive/2026/07)
- [June 2026 (2)](https://blog.seqgen.com/archive/2026/06)
- [March 2026 (2)](https://blog.seqgen.com/archive/2026/03)
- [February 2026 (2)](https://blog.seqgen.com/archive/2026/02)
- [January 2026 (4)](https://blog.seqgen.com/archive/2026/01)
- [December 2025 (3)](https://blog.seqgen.com/archive/2025/12)
- [November 2025 (1)](https://blog.seqgen.com/archive/2025/11)
- [October 2025 (4)](https://blog.seqgen.com/archive/2025/10)
- [September 2025 (5)](https://blog.seqgen.com/archive/2025/09)
- [August 2025 (1)](https://blog.seqgen.com/archive/2025/08)
- [July 2025 (1)](https://blog.seqgen.com/archive/2025/07)
- [June 2025 (2)](https://blog.seqgen.com/archive/2025/06)
- [May 2025 (1)](https://blog.seqgen.com/archive/2025/05)
- [April 2025 (2)](https://blog.seqgen.com/archive/2025/04)
- [March 2025 (2)](https://blog.seqgen.com/archive/2025/03)
- [February 2025 (2)](https://blog.seqgen.com/archive/2025/02)
- [July 2023 (1)](https://blog.seqgen.com/archive/2023/07)
- [May 2023 (1)](https://blog.seqgen.com/archive/2023/05)
- [April 2023 (2)](https://blog.seqgen.com/archive/2023/04)
- [March 2023 (3)](https://blog.seqgen.com/archive/2023/03)
- [October 2022 (1)](https://blog.seqgen.com/archive/2022/10)
- [September 2022 (1)](https://blog.seqgen.com/archive/2022/09)
- [May 2022 (1)](https://blog.seqgen.com/archive/2022/05)
- [February 2022 (3)](https://blog.seqgen.com/archive/2022/02)
- [February 2021 (1)](https://blog.seqgen.com/archive/2021/02)
- [March 2020 (2)](https://blog.seqgen.com/archive/2020/03)
- [December 2019 (1)](https://blog.seqgen.com/archive/2019/12)
- [November 2019 (1)](https://blog.seqgen.com/archive/2019/11)
- [September 2019 (1)](https://blog.seqgen.com/archive/2019/09)
- [June 2019 (1)](https://blog.seqgen.com/archive/2019/06)
- [August 2018 (1)](https://blog.seqgen.com/archive/2018/08)
- [May 2018 (1)](https://blog.seqgen.com/archive/2018/05)
- [April 2018 (1)](https://blog.seqgen.com/archive/2018/04)
- [March 2018 (1)](https://blog.seqgen.com/archive/2018/03)
- [January 2018 (1)](https://blog.seqgen.com/archive/2018/01)
- [December 2017 (1)](https://blog.seqgen.com/archive/2017/12)
- [November 2017 (1)](https://blog.seqgen.com/archive/2017/11)
- [October 2017 (1)](https://blog.seqgen.com/archive/2017/10)
- [September 2017 (1)](https://blog.seqgen.com/archive/2017/09)
- [August 2017 (2)](https://blog.seqgen.com/archive/2017/08)
- [July 2017 (2)](https://blog.seqgen.com/archive/2017/07)
- [June 2017 (2)](https://blog.seqgen.com/archive/2017/06)
- [May 2017 (5)](https://blog.seqgen.com/archive/2017/05)
- [April 2017 (1)](https://blog.seqgen.com/archive/2017/04)
- [March 2017 (5)](https://blog.seqgen.com/archive/2017/03)
- [February 2017 (4)](https://blog.seqgen.com/archive/2017/02)
- [January 2017 (4)](https://blog.seqgen.com/archive/2017/01)
- [December 2016 (3)](https://blog.seqgen.com/archive/2016/12)
- [November 2016 (3)](https://blog.seqgen.com/archive/2016/11)
- [October 2016 (4)](https://blog.seqgen.com/archive/2016/10)
- [September 2016 (3)](https://blog.seqgen.com/archive/2016/09)
- [August 2016 (5)](https://blog.seqgen.com/archive/2016/08)
- [July 2016 (3)](https://blog.seqgen.com/archive/2016/07)
- [June 2016 (3)](https://blog.seqgen.com/archive/2016/06)
- [May 2016 (4)](https://blog.seqgen.com/archive/2016/05)
- [April 2016 (4)](https://blog.seqgen.com/archive/2016/04)
- [March 2016 (4)](https://blog.seqgen.com/archive/2016/03)
- [February 2016 (30)](https://blog.seqgen.com/archive/2016/02)

see all

## Contact Us

#### SeqGen, Inc.

1725 Del Amo Blvd Torrance, CA 90501

#### +1 877 377 3743

#### +1 800 790 4830

#### [info@seqgen.com](mailto:%20info@seqgen.com)

#### Opening Hours

Monday - Friday: 8:30AM PST - 5:30PM PST

## Social Links

- <https://www.facebook.com/SeqGen/>
- <https://twitter.com/seqgen>
- <https://www.instagram.com/seqgen/>
- <https://www.linkedin.com/company/seqgen-inc.>

 SeqGen is an independently owned and operated provider of repair services and refurbished laboratory equipment. Any product names, logos, brands, and other trademarks or images featured or referred to within the SeqGen website and/or on any social media forum are the property of their respective trademark holders or OEMs. These trademark holders are not affiliated with SeqGen or its website. These trademark holders do not sponsor or endorse SeqGen or any of its products or comments. The use of any trade name or trademark is for identification and reference purposes only and does not imply any association with the trademark holder of their product brand.

```json
{
  "@context" : "https://schema.org",
  "@type" : "BlogPosting",
  "author" : {
    "@type" : "Person",
    "name" : "Service Team",
    "url" : "https://blog.seqgen.com/author/service-team"
  },
  "dateModified" : "2026-07-30T13:37:06.731Z",
  "datePublished" : "2025-09-10T13:00:00.000Z",
  "headline" : "Fix the Intro.ini Missing or Incorrect Error on Your Tecan EVOware",
  "image" : [ "https://blog.seqgen.com/hubfs/tecan-intro-is-missing.png" ],
  "mainEntityOfPage" : {
    "@id" : "https://blog.seqgen.com/fix-the-intro.ini-missing-or-incorrect-error-on-your-tecan-evoware",
    "@type" : "WebPage"
  },
  "publisher" : {
    "@type" : "Organization",
    "logo" : {
      "@type" : "ImageObject",
      "url" : "https://blog.seqgen.com/hubfs/SmallLogo-2.png"
    },
    "name" : "SeqGen, Inc."
  }
}
```