# [Troubleshooting the Te-Motion Error EVO\_EVO\_013\_230 on Your Tecan](https://blog.seqgen.com/troubleshooting-the-te-motion-error-evo_evo_013_230-on-your-tecan)

 Written by [Service Team](https://blog.seqgen.com/author/service-team) | 03 September 2025

Encountering an error message like "**EVO\_EVO\_013\_230: Te-Motion error**" on your channel/cable assembly can be frustrating, especially when it prevents you from using the instrument. This error often appears to happen randomly, sometimes during initialization or in the middle of a script, particularly when the system is aspirating, dispensing, or picking up a Diti tip. While a simple restart might clear the error temporarily, it often returns and eventually becomes permanent.

## What is a Te-Motion Error?

A Te-Motion error indicates an issue with the movement or control of a specific channel's Z-rod motor. Think of it as a signal that something is mechanically wrong with the channel assembly itself. It's often not a software glitch but rather a hardware malfunction. This is why restarting the software or instrument may provide a brief reprieve but won't solve the underlying problem.

## Common Symptoms

You'll likely see this error message in your software, but here are some of the actions that might trigger it:

- **Getting a Diti tip:** The channel struggles to pick up a new disposable tip properly.
- **Aspirating liquid:** The diluter's syringe or plunger isn't moving correctly to draw in the liquid.
- **Dispensing liquid:** The system fails to push the liquid out smoothly.

## The Solution: A Channel Assembly Replacement

Based on these symptoms, the most likely cause is an **Air Liha** **malfunctioning channel/cable assembly**. This assembly is a complex component essential for precise liquid handling. When it fails, the channel can no longer perform its movements reliably, leading to the Te-Motion error.

Since this is a hardware issue, it requires a physical replacement. Software resets and other troubleshooting steps won't work in the long term because they can't fix a broken mechanical part.

## What to Do Next

If you're experiencing a persistent "Te-Motion error" on a specific channel, the next step is to schedule a service to replace the **Air Liha Channel assembly**. This is the most effective way to resolve the issue and get your diluter back up and running.

Please [contact our service team](https://seqgen.com/service-request/) to schedule this replacement.

 

## Frequently Asked Questions (FAQs)

### 1. What is a Te-Motion error on a Tecan EVOware?

A Te-Motion error indicates an issue with the movement or control of a specific channel's Z-rod motor. It is a signal that something is mechanically wrong with the channel assembly itself — it is often not a software glitch, but rather a hardware malfunction.

### 2. When does the Te-Motion error EVO\_EVO\_013\_230 appear?

The error often appears to happen randomly, sometimes during initialization or in the middle of a script, particularly when the system is aspirating, dispensing, or picking up a Diti tip.

### 3. What are the common symptoms of a Te-Motion error?

The channel struggles to pick up a new disposable tip properly, the diluter's syringe or plunger does not move correctly to draw in liquid when aspirating, and the system fails to push liquid out smoothly when dispensing.

### 4. What causes the Te-Motion error, and can a software reset fix it?

The most likely cause is a malfunctioning Air Liha channel/cable assembly — a complex component essential for precise liquid handling. Because this is a hardware issue, it requires a physical replacement. Software resets and other troubleshooting steps will not work in the long term because they cannot fix a broken mechanical part.

### 5. What should I do if the Te-Motion error keeps returning?

A simple restart might clear the error temporarily, but it often returns and eventually becomes permanent. If you are experiencing a persistent Te-Motion error on a specific channel, the next step is to schedule a service to replace the Air Liha channel assembly — the most effective way to resolve the issue and get your diluter back up and running. Contact SeqGen's service team to schedule the replacement.

 

SeqGen specializes in the repair, maintenance, and refurbishment of essential laboratory instruments used in DNA sequencing and genetic research. Our expertise covers a wide range of equipment, including DNA sequencers like the [ABI 3730/3730XL](https://seqgen.com/instrument-model/dna-analyzer/abi-3730-3730xl/), [3500/3500XL](https://seqgen.com/instrument-model/dna-analyzer/abi-3500-3500xl/), SCIEX LC/MS mass spectrometers such as the [SCIEX 6500+](https://seqgen.com/instrument-model/lc-ms/api-6500/), [TECAN Freedom EVO](https://seqgen.com/instrument-model/liquid-handler/tecan-freedom-evo/) liquid handlers, real-time PCR systems, thermal cyclers, and microplate readers. With experience in both current and legacy models, SeqGen provides cost-effective solutions at significantly lower prices than leading competitors.

[View full post](https://blog.seqgen.com/troubleshooting-the-te-motion-error-evo_evo_013_230-on-your-tecan)

```json
{
  "@context" : "http://schema.org",
  "@type" : "BlogPosting",
  "author" : {
    "@type" : "Person",
    "name" : "Service Team"
  },
  "dateModified" : "2026-11-19T23:24:29.174Z",
  "datePublished" : "2025-09-03T13:00:00Z",
  "headline" : "Troubleshooting the Te-Motion Error EVO_EVO_013_230 on Your Tecan",
  "image" : {
    "@type" : "ImageObject",
    "height" : 384,
    "url" : "https://1939955.fs1.hubspotusercontent-na2.net/hubfs/1939955/EVO_EVO_013_230%20error.jpg",
    "width" : 512
  },
  "mainEntityOfPage" : "https://blog.seqgen.com/troubleshooting-the-te-motion-error-evo_evo_013_230-on-your-tecan",
  "publisher" : {
    "@type" : "Organization",
    "logo" : {
      "@type" : "ImageObject",
      "height" : 60,
      "url" : "/hs/hsstatic/content_shared_assets/static-1.4092/img/default-amp-logo.png",
      "width" : 60
    },
    "name" : "SeqGen Blog"
  }
}
```