But something was wrong. Two strands of lights decided to be their own light show, blinking and changing colors on their own command. Frustrated, I began to dig into the wiring, assessing everything from the bulb fuses to the circuitry. Nothing I did fixed the issue. Wouldn’t it be nice if Christmas lights came with troubleshooting instructions? Or, at the very least, error messages that pinpointed a general cause?
Your sequencing instrument, on the other hand, can display a variety of error messages. But typically, it’s up to you, the end user, to decipher the issue your instrument is trying to communicate.
Here are a few steps to follow if you ever encounter this message on an ABI 3500: “ERROR: leak detected during bubble compression.”
Often, the issue is the one-way valve. This is a small silver valve where the polymer tube connects to the pump upper block. If the one-way valve leaks, it will cause the polymer in the pump to flow back into the polymer pouch when you fill the array with polymer.