Troubleshooting Mass Spectrometers 4500/5500/6500

Having issues activating your Mass Spectrometer profile?

Our expert team has helped many users to activate the profile in the hardware configuration. Troubleshooting your Mass Spectrometer is essential to figure out any issue when you cannot activate it. 

Accidentally turning off ​the gas generator can cause a malfunction. The software won’t finish its configuration if the pressure of the curtain gas is too low. For this reason, the first thing to do is to check if the gas generator is working correctly. If the issue was the gas generator, make sure it has been turned on, and the gas pressure of curtain gas is higher than 60 psi. 

Reset the instrument. To reset your instrument, push the Reset Button located at the left corner of the left side panel. Hold the button for 5 seconds and then release. Furthermore, to complete the process, reboot your computer.

After its completion, this process should allow you to configure the profile in the hardware.

Please follow the instructions below if you determine that the issue was not the gas generator.

1. Use search to find Services App. and open it.

MS Services app

2. Stop and Re-start the AnalystService in the “services.”

3. Select AnalystService and click “stop” the service.

4. Re-start the service

5. Open Analyst and configure hardware again

6. If it is still not working, Use Command Prompt to ping the instrument.

7. Type cmd in search and open the command prompt

8. Ping 192.168.100.2

If it is unreachable, please check the connection. If the connection is good, then check the IP address setting. If the IP address setting is correct, the ICB board in the instrument or the network card in the computer may be broken.

For expert advice and help with troubleshooting, repair services, annual contracts, or preventative maintenance, reach out to sales@seqgen.com.

Written by Service Team

Subscribe to Email Updates

New Call-to-action 

Recent Posts

SeqGen is an independently owned and operated provider of repair services and refurbished laboratory equipment. Any product names, logos, brands, and other trademarks or images featured or referred to within the SeqGen website and/or on any social media forum are the property of their respective trademark holders or OEMs. These trademark holders are not affiliated with SeqGen or its website. These trademark holders do not sponsor or endorse SeqGen or any of its products or comments. The use of any trade name or trademark is for identification and reference purposes only and does not imply any association with the trademark holder of their product brand.